page 46 of 121
• Ensure that the
coolant outlet tubing
at both con-
densers is always unobstructed and that it cannot get
blocked.
• Water pressure shock at the condensers may cause
breakage of the condenser coil!
Notes concerning the operation of the Peltronic emis-
sion condenser
Read the manual of the Peltronic emission condenser and
observe the instructions contained in that manual!
Assembling the Peltronic emission condenser:
➨
Assemble the Peltronic emission condenser
with two attachment screws (Phillips screw-
driver size 2) to the pumping unit (a).
➨
Assemble the hose connection between
pump outlet and condenser inlet (b).
➨
Slip connecting tube onto hose connection
of the Peltronic emission condenser. Se-
cure with hose clip.
➨
Connect the VACUU•BUS cable from the
CVC 3000 controller to the Peltronic emis-
sion condenser (c).
Notes concerning the operation of pumping units with
silencer
Attention
: Dust-laden gases, deposits and condensed
solvent vapor can restrict air flow out the silencer. The re
-
sultant back pressure can lead to damage of pump bear-
ings, diaphragms, and valves. Under those conditions, a
silencer must not be used.
NOTICE
(a)
(b)
(c)
NOTICE