page 110 of 121
➨
Tighten the union nut (K) of the hose con-
nection with an open-ended wrench w/f 17.
+
Tighten the union nut first by hand and then
tighten one full turn using the open ended
wrench.
PC 3001 VARIO
pro
/ PC 3001 VARIO
pro
IK
Assemble the exhaust vapor condenser.
➨
Feed the PTFE hose from the pump outlet
into the inlet of the exhaust vapor condens-
er.
➨
Mount the condenser with the counter hold-
er (Torx screws T10) to the pump.
➨
Fasten union nut.
PC 3001 VARIO
pro
IK
Assemble the inlet condenser.
➨
Slip the hose from the inlet of the pump onto
the hose connector of the inlet condenser.
➨
Mount the condenser with the counter hold-
er (Torx screws T10) to the pump.
➨
Tighten the union nut first by hand and then
tighten one full turn using the open ended
wrench.
PC 3001 VARIO
pro
TE
Assemble the emission condenser TE.
➨
Feed the PTFE hose from the pump outlet
into the inlet of the emission condenser TE.
➨
Attach the mounting plate of the emission
condenser TE with 2 screws at the hous-
ing cover of the pumping unit; use a Phil-
lips screw driver size 2. Pay attention to the
cage nuts in the groove of the housing cov-
er.
➨
Fasten union nut.