page 101 of 121
Cleaning and inspecting the pump heads
➨
Remove catchpots (10) at inlet and outlet
(see „Use and operation“, pg. 38).
Detach the coupling of the hose connection
(M) below the housing cover (A).
➨
Open the hose clip (N) with a slotted screw-
driver.
➨
Pull the tubing off the hose connector.
Opening the hose clip:
➨
Apply slotted screwdriver as shown and
turn.
N
M
A
PC 3001 VARIO
pro
/ PC 3001 VARIO
pro
IK
Disassemble the exhaust vapor condenser.
➨
Loosen the union nut at the inlet of the ex
-
haust vapor condenser.
➨
Remove the 4 screws affixing the counter
holder (O) of the exhaust vapor condenser
with a Torx driver T10.
➨
Remove the exhaust vapor condenser. In
doing so pull the PTFE hose out of the inlet
of the exhaust vapor condenser.
O