Step 2:
Ensure the quartz O-ring seal is in place around the quartz sleeve
and then replace the top manifold by aligning the shaft adapter into the seal
cavity of the top manifold (see Figure 7.8).
Step 3:
Tighten the nuts in an alternating pattern. Once top manifold is
secure, replace the retaining ring on the wiper shaft. Rotate the wiper
assembly CCW to have the top wiper blade facing the back of the unit.
Step 4:
Reconnect both top and bottom piping connections. Close the UV
Chamber door.
Step 5:
Close any taps and slowly open the water supply. Inspect for leaks.
Repair any leaks if necessary before plugging in the unit.
Step 6:
Reinstall the motor mounting plate by using the trigger cam to center
the plate (center hole to be concentric with wiper shaft). With the top wiper
blade facing the back, the trigger cam should have the dimple in the 10
o’clock position. Install the wiper motor, then reconnect all the wires.
Step 7:
Reinstall the top and bottom plastic covers.
Step 8:
Plug in the unit. Check operation of wiper motor, purge valve and
water temperature switch.
Step 9:
Make an entry in the service record to establish a cleaning schedule.
Replacement Parts
Use only genuine parts from UV Pure Technologies when servicing your
Hallett disinfection system. Failure to use genuine replacement parts will void
the factory warranty, and any laboratory validation and/or certification for
water safety and system operating performance. Figures 7.9A-B shows a
complete list of original factory parts.
Replacement parts and service are available from your Installer, a Certified
Service Dealer (CSD), or directly from UV Pure Technologies on our
e-commerce site:
. A complete list of Certified Service
Dealers and their coverage areas is available on UV Pure’s website:
Manufactured by:
UV Pure Technologies Inc.
455 Milner Avenue, Unit 1
Toronto, Ontario, Canada, M1B 2K4
416-208-9884, 1-888-407-9997
FAX 416-208-5808
e-mail inquires:
www.uvpure.com
Date of Purchase:
Dealer Name:
Model:
Serial Number:
(S/N located on the left side of the front panel below power cord entry)
Содержание Crossfire Technology Hallett 1000NC
Страница 41: ...Hallett Instruction Manual Released on 8 Aug 19 GH2B Page 41 UV Pure Technologies Inc 2019 Figure 7 9A...
Страница 50: ...Hallett Instruction Manual Released on 8 Aug 19 GH2B Page 50 UV Pure Technologies Inc 2019...
Страница 51: ...Hallett Instruction Manual Released on 8 Aug 19 GH2B Page 51 UV Pure Technologies Inc 2019...
Страница 52: ...UV Pure Technologies Inc 455 Milner Ave Unit 1 Toronto Ontario M1B 2K4 www uvpure com...