Hallett Instruction Manual
Released on 8-Aug-19
– GH2B
Page 33
UV Pure Technologies Inc. © 2019
Step 3:
Install the new lamps into the unit, one at a time being careful not to
touch the bulb. Insert the top end of the lamp through the top lamp holder
then swing lamp base in over the socket. For a LPHO lamp, rotate it until the
“Stop Sign” symbol printed on the top ceramic is towards the left. See Figure
7.5. For an amalgam lamp, rotate it until the wires running down the lamp are
facing the back of the unit. The proper orientation of the LPHO and amalgam
lamp has the lamp wires opposite the UV sensors.
Figure 7.5
After the lamp has been rotated correctly, allow the 4 pins of the lamp to
engage into the socket, push firmly down on the top of the lamp.
Do not twist
the lamps when they are inserted.
Observe the base of the lamp to confirm
the pins are fully seated into the socket. Repeat for other lamp.
Step 4:
Close the UV Chamber and secure door.
Step 5:
Record the date of the lamp replacement in your Service Record
Sheet
.
Step 6:
Plug in the unit. The lamp lifetime counter can be reset in the
Settings menu, item 2.5 (365 days for LPHO lamps, 500 days for amalgam
lamps).
New LPHO lamps may take from a few moments to several hours
Содержание Crossfire Technology Hallett 1000NC
Страница 41: ...Hallett Instruction Manual Released on 8 Aug 19 GH2B Page 41 UV Pure Technologies Inc 2019 Figure 7 9A...
Страница 50: ...Hallett Instruction Manual Released on 8 Aug 19 GH2B Page 50 UV Pure Technologies Inc 2019...
Страница 51: ...Hallett Instruction Manual Released on 8 Aug 19 GH2B Page 51 UV Pure Technologies Inc 2019...
Страница 52: ...UV Pure Technologies Inc 455 Milner Ave Unit 1 Toronto Ontario M1B 2K4 www uvpure com...