Page
5
of
6
7. STORAGE:
Store the N
2
analyzer in a temperature similar to its ambient environment of daily use. The
instructions given below describe the methods to clean the instrument, sensor and its accessories:
8. CLEANING
Instrument
:
When cleaning or disinfecting the exterior of the N2 analyzer, take appropriate care to
prevent any solution from entering the instrument. Do not immerse instrument in fluids.
Oxygen Sensor
:
Clean the sensor with a cloth moistened with a 65% alcohol / water solution. Do
not use spray disinfectants because they can contain salt, which can accumulate in the sensor
membrane and impair readings.
Accessory
: The threaded barbed adapter may be cleaned by washing with a 65% alcohol / water
solution. The part must be thoroughly dry before it is re-used.
9. DISPOSAL
Do not throw away. Dispose of properly in accordance with local regulations. The sensor, batteries,
and circuit board are not suitable for regular trash disposal. Follow local guidelines for disposal of
components.
Содержание Nitrogen Pac SC-W
Страница 2: ... THIS SURFACE DELIBERATELY LEFT BLANK ...
Страница 81: ...Page 1 of 6 Model NA 2 Nitrogen Analyzer Hand Held OPERATING MANUAL INSTRUCTIONS FOR USE R218M06 Industrial ...
Страница 102: ...APPENDIX F SC W MANUAL ...
Страница 103: ...APPENDIX F SC W MANUAL ...
Страница 104: ...APPENDIX G SC W MANUAL ...
Страница 111: ... THIS SURFACE DELIBERATELY LEFT BLANK ...