Chapter
13
■
Continuous transmitting format
- Transmitting format 1
- Transmitting format 2
*: When analog peak hold (peak hold selection = 1: analog, and hold mode = 1: peak hold) is used,
the 5 digits are the same as the linked indicated values.
- Transmitting format 3
*1
O: Overload
(LOAD, OFL, OVERLOAD)
S: Stable
M: Unstable
H: Hold
Priority: H
>
O
>
(S or M)
*4
N: Near zero OFF
Z: Near zero ON
*5
H: HH limit ON
L: LL limit ON
G: HH limit and LL limit OFF
N: HH limit and LL limit ON
F: Compare OFF
Priority: N
>
(H or L) F
>
G
*2
A: Zero tracking OFF
T: Zero tracking ON
*3
H: HI limit ON
L: LO limit ON
G: HI limit and LO limit OFF
N: HI limit and LO limit ON
F: Compare OFF
Priority: N
>
(H or L) F
>
G
*6
Decimal place
3: 0.000
2: 00.00
1: 000.0
0: None
0 1 2
3 4 5
6 7
8 9 10 11 12 13 14 15 16 17 18 19
G S
,
*1
*2
*3
*4
,
,
,
,
Indicated value linked (5 decimal point)
Sign
+
or
-
±
*5 ,
20
21
Terminator
0 1 2
3 4 5
6 7
8 9 10 11 12 13 14 15 16 17 18 19 20 21
W T
*1 *2 *3 *4
±
Indicated value not
Sign
+
or
-
Indicated value
*5
SOH
STX
ETX BCC
Sign
+
or
-
±
*6
22 23
BCC calculation range
Each ASCII code of SOH, STX, ETX
SOH: 01
STX: 02
ETX: 03
BCC calculation method
For BCC,
each code in the BCC calculation range is represented in
hexadecimals and is XORed with all data.
linked (5 digits)
linked (5 digits)*
0 1 2
3 4 5
6 7
8 9 10 11 12 13 14 15 16 17 18
19
G S
,
*1
*2
*3
*4
,
,
,
Indicated value linked (5 decimal point)
Sign
+
or
-
±
,
Terminator
Содержание F325
Страница 1: ...01JUN2015REV 1 01 DIGITAL INDICATOR F325 OPERATION MANUAL...
Страница 153: ...142 15 Specifications 142 Specifications Chapter 15 When BCO BSC option is selected Unit mm When BSC option is selected...
Страница 154: ...143 15 Specifications 143 Specifications Chapter 15 When DAV DAI option is selected Unit mm When DAI option is selected...
Страница 155: ...144 15 Specifications 144 Specifications Chapter 15 When RS 232C option is selected Unit mm...
Страница 160: ......