Settings and Operations Related to Hold
53
Se
tting
s an
d Op
er
ations Related
to
Hold
Chapter
7
■
Peak hold (maintaining a maximum point)
When hold fix section is 0: OFF
●
Operation of peak hold
* Hold output is performed if a hold is set to output selection 1 or output selection 2.
HOLD input
Input
Time
Sensor input value
Indicated value
Peak hold section
HOLD LED
Off
On
Peak detection/hold
Peak hold reset
PEAK LED
Off
Off
Flashing
Off
HOLD
Off
On
Peak hold section
OFF
ON
Hold output
OFF
ON
OFF
ON
key
- When hold is turned ON with external signals, turn it OFF with external signals.
When hold is turned ON with key inputs, turn it OFF with key inputs.
- Refer to P.17 “2-8.External input connection” for connection of hold input
terminals.
Key points