D/A resolution
D/A converter resolution is 1/10000 to 4 to 20mA.
Minimum unit of current is
(20
-
4mA) x 1/10000 = 1.6
μ
A.
Minimum unit of indicated value is
(D/A full scale setting value - D/A zero setting value) x 1/10000.
■
Setting values related to D/A converter
■
D/A zero/full scale
D/A zero/full scale of F325 is set.
■
D/A zero/full scale setting method
1.
Select setting mode 4.
→
(three times)
2.
Set "D/A zero setting".
Press the
key five times.
Set the D/A zero setting using the
(numerical value input) and
(shift) keys, and
confirm it with the
key.
Press the
key to insert a minus sign.
3.
Set "D/A full scale setting".
Press the
key once.
Set the D/A full scale setting using the
(numerical value input) and
(shift) keys, and
confirm it with the
key.
Press the
key to insert a minus sign.
4.
Press the
key to return to the indicated value display.
FNC
+
-
ZERO
D/A zero value
(-19999 to 99999)
CAL
HOLD
GAIN
CAL
ZERO
HOLD
FNC
+
-
D/A full scale value
(-19999 to 99999)
CAL
HOLD
GAIN
CAL
ZERO
HOLD
FNC
+
-
FNC
+
-
Содержание F325
Страница 1: ...01JUN2015REV 1 01 DIGITAL INDICATOR F325 OPERATION MANUAL...
Страница 153: ...142 15 Specifications 142 Specifications Chapter 15 When BCO BSC option is selected Unit mm When BSC option is selected...
Страница 154: ...143 15 Specifications 143 Specifications Chapter 15 When DAV DAI option is selected Unit mm When DAI option is selected...
Страница 155: ...144 15 Specifications 144 Specifications Chapter 15 When RS 232C option is selected Unit mm...
Страница 160: ......