System Settings and Operations
67
Syste
m
Settin
gs and
Ope
ratio
ns
Chapter
9
9-2.
Password
By inputting a password, memory contents can be rewritten (initialized) to the factory default
contents.
Calibration values (zero calibration, span calibration) remain unchanged; however, other setting
values can be rewritten to the factory default values.
■
Password setting method
1.
Select setting mode 4.
→
(three times)
2.
Set "password".
Press the
key eight times.
Set the password using the
(numerical value input) and
(shift) keys, and
confirm it with the
key.
Initialization is complete once the tone sounds.
(Do not operate keys until the tone sounds.)
FNC
+
-
ZERO
Password
1239: Initialization
CAL
HOLD
GAIN
CAL
ZERO
HOLD
- Refer to P.146 “16.Setting Values List” for initialized values.
- Refer to P.68 “9-3.Self-check/Initialization” if self-check and initialization are to
be simultaneously performed.