41
4
SETTINGS AND CALIBRATIONS
41
SETTINGS AND CAL
IBRATIONS
Chapter
4
6.
Overload
Set the load at which the sensor is judged to be overloaded. The
Load OK output turns OFF when the load exceeds the set value.
Operation
Setting range (0 to 9999)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Overload
7.
Increment
Set the minimum unit of load display. Loads are displayed in the
set increments.
Operation
Setting range (1, 2, 5, 10)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Increment
8.
Analog Filter
The low-pass filter of the set cut-off frequency is inserted with
respect to the input of the load sensor.
Operation
Setting range (10, 30, 100, 300 Hz)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Next
→
Analog Filter
9.
Digital Filter
Loads are moving-averaged by the set number of times.
Operation
Setting range (0, 2 to 999 Times)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Next
→
Digital filter
10.
Digital Zero Limit
Set the limit of load that can be digital-zeroed. If Digital Zero is
performed with a load larger (smaller) than the zero-calibrated
load by the set load, an error will result, and the load will not be
zeroed.
Operation
Setting range (0 to 9999)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Next
→
DZ Limit
Содержание DeviceNet F381A
Страница 1: ...15APR2013REV 3 10 DYNAMIC FORCE PROCESSOR F381A OPERATION MANUAL ...
Страница 9: ...Contents VIII Contents VIII M E M O ...
Страница 34: ...25 2 INSTALLATION AND CONNECTION 25 INSTALLATION AND CONNECTION Chapter 2 M E M O ...
Страница 147: ...138 8 SPECIFICATIONS 138 SPECIFICATIONS Chapter 8 8 2 Outside dimensions Unit mm ...
Страница 164: ......