UT4000 User Manual
Chapter 6
⎯
Appendixes
Appendix A : Technical Indicators
Unless otherwise specified, all technical specifications apply to probes with a attenuation
setting of 10X and the UT4000 Series Digital Oscilloscopes. To verify that your
oscilloscope meets specifications, it must first meet the following conditions :
The oscilloscope must have been operating continuously for thirty minutes within
the specified operating temperature.
If the operating temperature changes by more than 5°C, you must perform the “Self
Cal” operation, accessible through the System Functions menu. All specifications
are guaranteed unless noted “Typical”.
Technical Indicators
Acquisition
Acquisition modes
Real time
Equivalent
Acquisition rates
2GS/s
50GS/s
(UT4302C/UT4202C/UT4152C)
Average
When all channels have made N acquisitions simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 to 256.
Input
Input coupling
DC, AC, GND
Input impedance
1M
Ω
±1.5% in parallel with 24pF±3pF.
Probe attenuation factor
1X, 10X, 100X, 1000X.
Maximum input voltage
400V (DC + AC Peak, 1 M
Ω
input impedance)
(10:1 probe attenuation); 5V (Vrms, BNC).
Time delay between channels (Typical) 150ps
Содержание UT4000 Series
Страница 13: ...2 UNI T UT4000 User Manual Figure 1 1 Front panel of UT4000 Oscilloscopes ...
Страница 14: ...3 UNI T UT4000 User Manual Figure 1 2 Schematic diagram for operating the UT4000 front panel ...
Страница 15: ...4 UNI T UT4000 User Manual Figure 1 3 Schematic diagram for the display interface ...
Страница 74: ...63 Figure 3 14 Dual time base function ...
Страница 98: ...87 UNI T UT4000 User Manual This user manual may be revised without prior notice ...