Using the Run Button
There is a [RUN/STOP] button on the top right corner of the front panel. When this
button is pressed and a green indicator lights up, your oscilloscope is in a running status.
When the red light comes on after pressing this button again, it indicates the unit has
stopped operation.
Figure 2-28 RUN/STOP button
Auto Setup :
As shown above, Auto Setup can simplify operation. Press [AUTO] and the oscilloscope
can automatically adjust the vertical deflection factor and horizontal time base range
according to the amplitude and frequency of the waveform, so as to ensure stable
waveform display. When the oscilloscope is in auto setup mode, the system setup is as
shown by Table 2-24 :
Table 2-24
Auto setup
Function Menu
Setup
Acquisition mode
Adjust to “sampling” or “peak detect”.
Display format
Set to YT.
SEC/DIV
Adjusted according to signal frequency.
Trigger coupling
AC
Trigger level
Set at 50%.
Trigger mode
Auto
Trigger source
Set to CH1 but if there is no signal in CH1 and CH2 applies
a signal, it will set to CH2.
Trigger inclination
Rising
Trigger type
Edge
Vertical bandwidth
Full
VOLT/DIV
Adjusted according to amplitude of the signal.
Содержание UT4000 Series
Страница 13: ...2 UNI T UT4000 User Manual Figure 1 1 Front panel of UT4000 Oscilloscopes ...
Страница 14: ...3 UNI T UT4000 User Manual Figure 1 2 Schematic diagram for operating the UT4000 front panel ...
Страница 15: ...4 UNI T UT4000 User Manual Figure 1 3 Schematic diagram for the display interface ...
Страница 74: ...63 Figure 3 14 Dual time base function ...
Страница 98: ...87 UNI T UT4000 User Manual This user manual may be revised without prior notice ...