7
UNI-T
®
UT4000 User Manual
d
You have to set the probe attenuation factor of the oscilloscope. This factor changes the
vertical range multiple to ensure the measurement result correctly reflects the amplitude
of the signal being tested. Set the attenuation factor of the probe as follows : Press [F4]
then [F2] to display 10X on the menu.
Figure 1-6
Setting the probe deflection factor of the oscilloscope
e
Connect the probe tip and ground clamp to the corresponding probe compensation
signal terminals. Press [AUTO] and you will see a square wave in the display (about 3V
peak-to-peak at 1kHz) in a few seconds, as shown in Figure 1-7.
4
Repeat these steps to check CH2. Press the [OFF] function button to close CH1, then
press [CH2] function button to open CH2. Repeat steps 2 and 3.
Figure 1-7
Probe compensation signal
Probe ration