81
/
99
Figure 13-7
Click the confirmation after the input is completed, it can continuous to access according to the new
modified frp proxy IP address information (in correct configuration)
Note:
If each oscilloscope is connect with the same frp server, then web_port
,
pic_port
,
ctrl_port of each
oscilloscope should be consistent , otherwise frp proxy fails and cannot access.
If frp proxy is modified, it cannot access by local area network IP: 9000. If user need to recover the
normal access, press the DEFAULT button on the oscilloscope panel to reset the configuration and then
use IP: 9000 port to access.
(4) Password Setting
Password setting is for the user account to log in. After the user changes the login password, the
user needs to use the new password for the next login. If you forget the password, reset the
password by pressing the DEFAULT button on the oscilloscope panel.