SonixTablet Service Manual
00.053.129, Revision C
Chapter 12: Troubleshooting Issues
12-3
12.7 IMAGE IS GRAINY OR MISSING
Check to ensure
Clarity
is set to
Medium
or
High
:
•
A.4.1 Adjust the Clarity Setting
.
Check to ensure the
Clarity License
is
Enabled
:
•
A.4.2 Check the Clarity License Status
.
Re-import
license.txt
to ensure that
Licenses
are not set to
Disabled
:
•
.
Check to ensure the transducer is not defective:
• test the transducer on another system and compare images
• submit test images for diagnosis to Ultrasonix Technical Support.
12.8 IMAGE HAS BLACK OR DARK AREAS
Black or dark areas on an ultrasound image are usually caused by dead elements in or delamination of the transducer. The
problem may be caused by either the transducer or the transducer MUX board, so both may have to be tested.
•
•
7.4.2 Testing the Transducer Board
Figure 12-1: Dead Element or Delamination Example
Содержание SonixTablet
Страница 1: ...SonixTablet Ultrasound System Service Manual...
Страница 2: ......
Страница 4: ......
Страница 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Страница 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Страница 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Страница 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Страница 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Страница 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Страница 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Страница 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Страница 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...