SonixTablet Service Manual
00.053.129, Revision C
Chapter 11: Maintenance
11-5
11.2.7.1 Cleaning the System Filter
To Clean the System Filter:
1.
Power off and unplug the system.
2.
Using a soft towel or bubble wrap to protect the LCD display/touch screen, gently tip the system flat.
3.
Unscrew the two (2) thumbscrews and washers.
4.
Push the filter forward (following the directional arrow label) then lift up.
5.
Vacuum thoroughly and
gently
reinstall the filter.
6.
Plug in and power on the system.
Note:
For additional views, refer to Service Drawing:
System Filter Removal/Replacement
.
Note:
Ensure the surface used is steady and secure.
Slide
Содержание SonixTablet
Страница 1: ...SonixTablet Ultrasound System Service Manual...
Страница 2: ......
Страница 4: ......
Страница 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Страница 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Страница 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Страница 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Страница 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Страница 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Страница 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Страница 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Страница 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...