Page 8 of 56 Pages
pressure gauge indicates to the operator that there is pressure in the
chamber.
A deviation of +1.6% is accepted.
The pressure scale, printer option can be set up at any time by a technician.
NOTE:
After operating the sterilizer, brown stains might appear on the bottom of
the chamber. These stains are a result of the heating elements that are
located at the lower external part of the chamber. The brown color is a
common phenomenon, can easily be removed, and will not have any effect
on the sterilized goods.
This manual is intended to give the user a general understanding of how
the autoclave works and indicates the best ways to operate and take care of
it in order to obtain optimum results and a trouble-free operation. After
reading this manual, operating the autoclave should be straightforward.
However, since the autoclave is built using high technology sensitive
components, no attempt should be made by the user or any other
unauthorized person to repair or recalibrate it.
Only technical personnel having proper qualifications, holding
technical documentation and adequate test instrumentation are
authorized to undertake repair or service.
2.2
Operating Conditions
Only autoclavable materials shall be used.
The ambient temperature shall be 5
º
C to 40
º
C and a relative
humidity respectively up to 85%
Operate the autoclave only in the manner specified in the manual. If
the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
Cautions!
Waste water should be brought into the public net in accordance
with the local rules or requirements i.e.
ONLY NON-HAZARDOUS LIQUIDS SHALL BE DISPOSED IN
PUBLIC SEWAGE!
The sterilizer should be loaded only with autoclavable material!
The environment shall not exceed an ambient temperature of 40
º
C
and a relative humidity of 85% respectively.
The operation altitude shall not be over 2000 meters (6561 feet)
(ambient pressure shall not be lower than 80 kPa (11.6 psi)).
The autoclave shall not be used in a manner not specified in this
manual!
Do not use the autoclave in the presence of dangerous gases.
The packed or unpacked device shall be stored in ‘indoor’ conditions.