Page 54 of 56 Pages
"Heat Time
Error" is
displayed.
The system cannot reach the required
temperature, in the chamber, within the
preset time.
Check that the autoclave is not
overloaded.
"Low
Pressure" is
displayed.
Pressure drops below the sterilization
pressure (134°C = 304 kPa, 121°C = 205
kPa) for 2 seconds during the sterilization
stage.
Perform a new cycle.
"High
Pressure" is
displayed.
Chamber Pressure raised 4.2 psi-29 kPa
above sterilization pressure (134°C = 304
kPa, 121°C = 205 kPa) for 2 seconds during
the sterilization stage.
Perform a new cycle.
"Pressure Time
Error" is
displayed.
The system cannot reach the required
pressure conditions in the chamber, after
preset time, during the air removal stage.
Check that the autoclave is not
overloaded.
"RTC Error -
Please Set
Current Date
and Time" is
displayed.
Date and the time are not set.
Set Current Date And Time.
If the problem persists, call the
technician.
"Time Error"
is displayed.
The real time clock is faulty.
Call the technician.
"Door is open
(During the
cycle)" is
displayed.
The door is open: During the cycle.
Close the door to perform a new
cycle.
"Canceled By
User" is
displayed.
The START/STOP key is pressed and cycle
aborted.
Wait until "cycle failed
–
canceled by user" or "cycle end
–
canceled by user" is displayed.
Perform a new cycle.
"Cycle Failed"
and the
following
symbol are
displayed.
An error occurred before sterilization cycle
is completed.
Perform a new cycle.
"Air Error" is
displayed at
the end of the
cycle.
The autoclave does not reach the
atmospheric pressure after 10 minutes.
Wait until the autoclave reaches
the atmospheric pressure and
perform a new cycle.
I/O Card
Failed
This message is displayed if I/O card is
faulty or disconnected (both while cycle is
running or not).
Call the technician.
"Power
Down" is
displayed.
power down occurred during the cycle. (this
message will be printed by the printer after
the autoclave will turn on).
Turn on the autoclave and wait
until the autoclave is ready
(reaches the safe condition) and
perform a new cycle.
If the problem is not covered by this section or if the problem persists,
contact your dealer or point of purchase for further assistance.