Page 50 of 56 Pages
Do not touch the strainer’s cover, mounted on the exhaust line, during
and short after operation.
Touching the hot strainer’s cover may cause severe injuries.
9.3
Checking the Safety Valve
Both safety valves, the chamber safety valve (2) and the generator safety
valve (3), are located on the top rear of the autoclave (see Rear View, 14
and 15).
They are protected by a safety shield.
In order to prevent the safety valve from a blockage, operate it once a
month.
Attention:
Use protective gloves in order to prevent hot steam burns to hands
9.3.1
PED-approved type safety valve
1.
For the autoclave safety valve, operate the sterilization
cycle; for the generator safety valve, operate the generator.
2.
Allow a pressure of approximately 200 kPa (29 psi) to
build up in the chamber.
3.
Turn the safety valve pressure-regulating nut clockwise
for 2 seconds.
Caution: Be careful not to burn your hands.
4.
Return the nut to its original position.
5.
Press the STOP/STOP key to interrupt the operation, and
exhaust steam from chamber.
6.
Wait until the pressure decreases to zero, only then can the
door be opened.
Rear Wall
Pressure-Regulating Nut