Page 5 of 56 Pages
If the autoclave is equipped with a printer, send along a copy of the last
printout for our inspection. If replacement parts are needed, stipulate the
model and serial number of the machine.
No autoclaves will be accepted for repair without proper authorization
from us. All transportation charges must be paid both ways by the owner.
This warranty will be void if the unit is not purchased from an authorized
full service Tuttnauer dealer.
1.4
Safety Instructions
The autoclave has unique characteristics. Please read and understand the operation
instructions before first operation of the autoclave. The following issues may
require guidance provided by the manufacturer: operation of the autoclave, the
door safety mechanism, the dangers involved in circumventing safety means, how
to ensure that the door is closed, and how to select a correct sterilization program.
Make sure that you know where the main power switch is, where the water cut-off
valve is and where the steam and compressed air disconnection valves are located.
Protective equipment and clothes and other safety instructions should be
implemented in accordance with local and national regulations and/or rules!
Autoclave maintenance is crucial for the correct and efficient function of the
device. We enclose a log booklet that includes maintenance recommendations,
with every device.
The daily B&D test and the weekly spore test is part of the preventive
maintenance plan, along with the annual validation of the sterilization processes
that ensures appropriate temperature dispersion within the chamber.
Never use the autoclave to sterilize corrosive products, such as: acids, bases and
phenols, volatile compounds or solutions such ethanol, methanol or chloroform
nor radioactive substances.
1.5
Operating Safety
1.
Never start using a new autoclave or a new steam generator, before the
safety, licensing and authorization department has approved it for use.
2.
All autoclave users must receive training in proper usage from an
experienced employee. Every new employee must undergo a training
period under an experienced employee.
3.
A written procedure must be established for autoclave operation,
including: daily safety tests, seal inspection and door hinge inspection,
smooth action of the closing mechanism, chamber cleaning, prevention of
clogging and preservation from corrosion, what is permitted and what is
prohibited for sterilization and choosing a sterilization program.
4.
Clean the steam generator daily.
5.
Before use, check inside the autoclave chamber to ensure that no items
have been left from the previous cycle.
6.
Load trays in such a way as to allow steam to move freely among all
items.