Page 6 of 56 Pages
7.
Only autoclavable products shall be used. Please refer to the manufacturer
instructions for sterilization of unknown materials or instruments.
8.
When sterilizing plastic materials, make sure that the item can withstand
sterilization temperature. Plastic that melts in the chamber is liable to
cause a great deal of damage.
9.
On closing the autoclave's door, make sure it is properly locked before
activating.
10.
Verify once again that you have chosen the appropriate sterilization
program.
11.
Before withdrawing trays, wear heat resistant gloves.
12.
Before opening the door, verify that there is no pressure in the chamber
(chamber pressure gauge is located on the autoclave's front panel).
13.
Open the door slowly to allow steam to escape and wait 5 minutes before
you remove the load.
14.
Once a month, ensure that the safety valves are functioning, and once
annually a certified tester must conduct pressure chamber safety tests.
15.
Once annually, or more frequently, effective tests must be performed, i.e.,
calibration and validation.
16.
Examine the condition of assemblies on a regular basis. Make sure there
are no leaks, breaks, blockages, whistles or strange noises.
17.
It is required to conduct maintenance operations as instructed.
18.
Immediately notify the person in charge of any deviation or risk for the
proper function of the device.
19.
This autoclave equipped with rear and front wheels, when moving the
autoclave all utilities must be disconnected, at least 2 persons shall move
the autoclave with caution. When the autoclave located in the required
place the Wheels has to be locked for further information see
FRONT/REAR VIEW OF THE AUTOCLAVE.
1.6
Safety features
Steam generator located inside the autoclave has the following
features for hazard protection:
1. Cut-off pressure switch
2. Cut-off thermostat
3. Water level protection
The main power of the generator is shut off if there is leak of water
or the pressure is above the safe level.
The pressure vessel chamber door has the following features
protecting personnel from hazards
1. Two door switches that indicate that the door is closed and
locked. Without this indication steam is not introduced to the
chamber.
2. A control that blocks door opening if the pressure is higher than
20% above the ambient pressure.