Page 7 of 56 Pages
2
TECHNICAL DATA
2.1
Introduction
This Horizontal autoclave is designed for the sterilization of wrapped and
unwrapped instruments and related items found in dental, medical and
veterinary clinics, first aid rooms, hospitals, laboratories, etc.
The instrument is a steam-heated sterilizer, using steam as sterilizing agent.
Its control unit ensures a fully automatic sterilization cycle, precise control and
monitoring of physical parameters and clear documentation of the sterilization
cycle. Five automatic programs are available, according to the material to be
ster two test programs.
`
This autoclave is equipped with a vacuum system and steam generator..
The advantages of the prevacuum sterilizer in comparison to the regular
gravity displacement steam sterilizer are as follows:
—
Removal of air pockets from packs and porous load and most kinds of
tubes (rubber, plastic etc.) by vacuum at the first stage of the cycle.
—
Better steam penetration into the load; resulting in effective
sterilization.
—
Better temperature uniformity.
This manual is intended to give the user a general understanding of how the
instrument works and indicate the best ways to operate and take care of it in
order to obtain optimum results and a trouble-free operation. After reading this
manual, operating the autoclave will be an easy job. However, since the
instrument is built using high technology sensitive components, no attempt
should be made by the user or any other unauthorized person to repair or re-
calibrate it.
Only technical personnel having the proper qualifications, and holding technical
documentation and adequate instrumentation are authorized to service the
instruments.
The temperature and pressure are controlled through sensors placed inside
and outside the media container.
The sterilizer has multiple built - in safety devices, which provides
adequate protection to ensure the safety of operating personnel.
On all models, a printer is an optional addition to the autoclave. The
printer prints the preset and actual parameters of the cycle (temperature,
time and pressure/vacuum).
The autoclave is provided with a pressure gauge that is used as guide only.
Should there be a power failure during the operation of the autoclave, the
Model 3870 HSG-D are table-top sterilizers designedespecially for the
sterilization of instruments, and other materials inhospital
laboratories, medical laboratories, research institutes, foodlaboratories and
pharmaceutical facilities.
A computerized control unit ensuring a fully automatic sterilization cycle
controls the autoclave.