Page 12 of 56 Pages
2.7
Autoclave's Electrical data
PROPERTIES
380-400V
3PH version
Ampere (A)
15 each phase
Voltage (V)
400/3ph
Hertz (Hz)
50/60
Heaters power (W)
9000
Protection against electrical shock
Class I
Degree of protection by enclosure
X0
2.8
Utilities
Water inlet (to generator)
1.5-2.5 Bar
Water connection(to generator)
1/2 NPT
Steam outlet
1/2 NPT
Drain
2" or bigger Withstanding
temp. of 80°C
Attention:
Attention:
The electrical net must be protected with a current leakage safety relay.
The electrical network must comply with local rules or regulations.
A circuit-breaker of 3 ph. 16A must be included in the building
installation. This switch or circuit-breaker shall be in close proximity
to the equipment, within easy reach of the operator; and marked as the
disconnecting device for the equipment.
The electrical net must be protected with a current leakage safety
relay.
The electrical network must comply with local rules or regulations.
Verify that there is an easy access to the main power switch, to the
water cut-off valve and to the current leakage safety relay.
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
All water connections to autoclave must be performed through
"BACK FLOW PREVENTION SYSTEM" only, as per IEC 61770.
The cable cord has to fit the electrical requirement rating.
Only locally qualified electrician can connect the autoclave to the
power.