MAN205-0097044ENO Rev. A
Apr. 2014
Page 14 of 28 pages
Note:
Verify that the packaging method is in accordance with good practice approach
and the packaging materials are in accordance with the applicable standards (e.g.
EN868 series).
4.1 Loading
Note:
Liquids can only be sterilized in an autoclave with two temperature sensors in the
chamber.
Every machine is supplied either with two (or more) stainless steel shelves, or with rails
for loading a loading cart from transfer carriage.
The transfer carriage is made of stainless steel and comes with adjustable legs to
balance the height of the loading cart to rails in the sterilizer chamber. The cart is
designed to slide smoothly off the carriage into the sterilization chamber.
The carriage moves on four heavy-duty castors, which are mounted on its base. The two
castors on the rear of the carriage are equipped with brakes. When the transfer carriage
is put in position for loading or un-loading the cart from the autoclave, the brakes must
be applied.
Caution!
It is strictly forbidden to load or unload the loading cart, if the transfer carriage
wheels brakes are not applied.
When loading the materials to be sterilized into the chamber, observe the rules
described in this chapter.
Do not overload the cart or the shelves as this can have adverse effects on the results of
the sterilization and drying. Leave space between the packs or wrapped instruments to
allow complete removal of air pockets in the air removal stage, and for free penetration
of steam in the heating and sterilization stages.
Transfer Carriage
Loading Cart