MAN205-0097044ENO Rev. A
Apr. 2014
Page 24 of 28 pages
7.
BACSOFT ERROR MESSAGES
7.1
BacSoft Touch Screen Stand by Error Messages
The following table describes the system error messages and corrective actions.
Message
Description
Corrective Action
System Ready
No error
For information only
Water in chamber
Water detected in the chamber
Call autoclave Technician
Door Is Open
Door is open
For information only
Cycle by clock is
active
Stand by cycle is scheduled to be run
NA
Jacket is cool
Jacket is cool
Wait till jacket reaches working
temperature
Steam generator
low pressure
Steam generator pressure is too low
Wait 30 minutes. (Applicable for
autoclaves with internal steam
generators
Drain is hot
Drain is too hot
For information only
IO Extension
device not
connected
Io Extension Not Connected (if IO
Extension is defined)
Call for service
High Start
Temperature
Accessory Cycle High Start
Temperature
Vhp Input Not
Ready
Accesory Cycle VhpIn Not Read (value
Call for service- to check the
vhp gen
Bio filter timeout
Unable to heat bio filter
Call for service to replace filter
Bio filter is cool
Thermo bio filter is too cold
Wait 20 minutes. If the
message not disappear call for
service
Printer port is not
valid
Printer port is not valid
Call for service
Wait...DoorOpening Door is opening
For information only
Wait…Door Closing Door is closing
For information only
Cool drain time out Time out while trying to cool drain
Call for service
Jacket time out
Time out while trying to heat jacket
Call for service
Supply error
No supply is detected
Call for service
Supply water error
No tap water supply
Call for service
Supply distilled
water error
No distilled water supply
Call for service