MAN205-0097044ENO Rev. A
Apr. 2014
Page 27 of 28 pages
Message
Description
Corrective Action
Heat Time Error
If chamber temperature does not
reach X°c as defined in cycle
parameter 'Sterilization
Temperature' after X minutes as
defined in cycle parameter 'Heat
Time Error'
Check utilities
or
Call for service
Low Pressure
If during sterilization stage, the
chamber pressure goes below the
minimum valid sterilization pressure
Check utilities
or
Call for service
High Pressure
If during sterilization stage, the
chamber pressure goes above the
maximum valid sterilization
pressure.
Check utilities
or
Call for service
Low Temp
If during sterilization stage, the
chamber temperature goes below
the sterilization temperature as
defined in cycle parameter
'Sterilization Temperature'
Check utilities
or
Call for service
High Temp
If during sterilization stage, the
chamber temperature goes above
the maximum valid sterilization
temperature
Call for service
Time Error
If there is a gap in sterilization time
of more than 3 seconds between the
two clocks of the machine
Call for service
Low Pressure
(Cooling)
If chamber pressure does not reach
X kPa as defined in cycle parameter
'Cool Pressure' after X minutes as
defined in cycle parameter 'Cool
Pressure Time Error'
Check and fix the compressed
air supply.
High Temp. (Cooling)
If chamber temperature does not
goes below X°c as defined in cycle
parameter 'Cool End Temperature'
after X minutes as defined in cycle
parameter 'Cool Error'
Call for service
High Pressure
(Exhaust)
If chamber pressure does not goes
below atmospheric pr 10
kPa after x minutes as defined in
cycle parameter 'Exhaust Time
Error'
Perform a new cycle.
High Pressure (Dry)
If chamber pressure is above
atmospheric pr 10 kPa
when staring dry stage
Call for service