14. At the end of the fragmentaTon procedure the
machine will stop to allow to control if the
fragmentaTon has been efficient (fig. 31). Open the
lid to control if the granules are present inside the
maker. If all is correct close the lid and push the
central buWon to end the procedure.
In this phase it’s possible to control the mill. You
can remove the tank and the mill to control inside.
Aqer this you can, following the starTng procedure,
close the lid and push the central buWon to start.
15. At the end of the procedure (about 20 min.), its
compleTon will be marked by an audio signal and
“END TREATMENT“ will appear on the display.
16. Open the door and remove the MAKER drum
carefully (fig. 32).
17. Place the MAKER drum where later it will be
used for surgery.
18. Carefully take out the CYLINDER which by this
stage will contain the liquids employed. Dispose
of the CYLINDER as special waste (fig. 33).
19. Carefully remove the liquids cartridge (TANK)
and dispose of it as special waste.
20. Carefully remove the MILL and treat it as any
contaminated instrument following standard
sterilizaTon procedures (fig. 34).
21. Turn off TT TRANSFORMER (fig. 35) and
disconnect power supply (fig. 36).
22. Clean the inside of the device using a damp
cloth and a cloth to dry (fig. 37).
Fig.31
Fig.32
Fig.34
Fig.35
Fig.36
Fig.37
Fig.33