Rev. 5 of 01.08.2020
16
4. USE OF TT TRANSFORMER
WARNING
: verify that all components are clean before using the device.
WARNING
: use transparent disposable tape (10,2 cm x 15,2 cm) on the device in order to protect
surfaces, commands and handle.
4.1 TOOTH PREPARATION
Aqer the extracTon, and before using the device, the tooth of the paTent must:
1. be cleaned from residual tartar using piezoelectric instrument;
2. have radicular surface cleaned using a diamond cuWer;
3. if needed, be cleaned from tooth’s enamel using a cuWer;
4. if possible, be broken in small parts to facilitate the work of the mill.
4.2 TOOTH INSERTION
The tooth of the paTent must be put inside the mill.
So it is necessary to:
1. open the Mill component and put inside the tooth;
2. close the Mill component.
4.3 START OF THE PROCESS
Connect the power supply to the main and turn ON the device using the general switch situated on the
back of the equipment.
When the word “WAIT” disappears on the display, open the lid and insert all components inside the
device according to the order described in the previous paragraph.
When all components are inserted and the door of the device is closed, aqer the display shows the
word “READY” it will be possible to press the “START” buWon to iniTate the process.
WARNING
: connect the power supply cord only to main that complies with all naTonal laws and
with an adequate earth protecTon (max 240Vac).
WARNING
: the device could be connected only to main with the features: 100-240Vac, 50-60Hz.
WARNING
: equipment has to be properly posiToned in order to facilitate the disconnecTon from the supply.