5. INSTRUCTIONS FOR
USE
1. Wear gloves and protecTve mask throughout the
procedure.
2. Apply medical adhesive tape on contact points.
3. Gently remove any residue of ligaments on the root
of the extracted tooth by using a diamond burr (fig.
25).
4. Remove any calculus and plaque residue from the
enTre tooth surface (fig.25).
5. Remove any reconstrucTon or foreign material
using a burr (fig.25).
6. Blow air across the surface of the remaining tooth
to dry it. The Tooth need to be perfectly dry.
7. Break the tooth up into liWle pieces.
8. Place the tooth pieces into the (perfectly dry) mill
(fig. 26).
9. Turn ON TT TRANSFORMER using the switch at the
back of the device (fig. 27).
10. Open the lid of TT TRANSFORMER when the word
WAIT disappears on the display (fig.28).
11. Insert the MILL, TANK , MAKER and CYLINDER
correctly into their respecTve slots (fig. 29).
12. Follow the pre-holes and pierce with an
instrument (fig.20).
13. Close the lid of TT TRANSFORMER (fig. 30).
14. Press the central buWon to start processing when
the word READY appears on the display (fig.30).
Fig.25
Fig.26
Fig.27
Fig.28
Fig.30
Fig.29
Fig.30