TruSteel AutoVap30 User Manual | | v. 1.9
26 / 39
WARNING All hoses may contain alcohol under pressure. Wear proper personal protective
equipment such as safety goggles and gloves when handling any part of the machine and
working around alcohol.
WARNING Component wear varies depending on the amount of usage. Inspect all nuts, bolts,
and gaskets before each use. If there is any question about the integrity of a component, replace
it immediately.
CAUTION The facility must provide adequate ventilation/exhaust, as determined by the Engineer
of Record, to maintain the local atmosphere below 25 of the Lower Flammability Limit (LFL).
CAUTION Alcohol vapor is heavier than air and can settle in low places.
CAUTION *DO NOT* use or store equipment or containers where they could be exposed to high
temperatures. Relief valves can open allowing alcohol to escape. DO NOT artificially heat
equipment or containers above 125°F.
CAUTION This system is NOT equipped with an overfill prevention device. Ensure proper fill
levels prior to operation.
CAUTION All system containers must be checked for accidental pressurization. Depressurize
prior to opening.
CAUTION *DO NOT* allow children or non-trained operators to tamper with or play near the
equipment.
HMI OPERATION
Four Sections of AV30 Control Panel Operation
1. Pump Control: Feed, Residue, and Ethanol Pump Auto/Manual Control and Setpoints
2. Temperature: Thermocouple temperature status for all TC
3. Alarms: Enable/Disable Alarms and adjust alarm setpoints
4. Alarm Reset: Resets Present Alarm
PUMP CONTROL
Pump Control Overview
·
All pumps have a general overview status on screen that will indicate speed being sent
to VFD drive in Hz
·
H-O-A Feedback Auto/Not in Auto
·
Pump Run Feedback Running/Stopped
·
Start/Stop control Push buttons
SYSTEM MODES
Extraction Mode
In extraction mode the system will automatically run all three pumps, and allow the Condenser
and Evaporator level to be maintained automatically. The Recycle Valve will also position itself
upon change of mode. There is currently a delay on the feed pump start to allow sufficient time
for the valve to re-position. When the motion indicator is on you will not be able to start the pump.
Recycle Mode
Содержание AutoVap30
Страница 1: ...Copyright 2020 by TruSteel LLC All Rights Reserved TruSteel AutoVap30 User Manual Version 1 9 ...
Страница 3: ...TruSteel AutoVap30 User Manual v 1 9 3 39 STEP 8 SHUT DOWN 39 ...
Страница 5: ...TruSteel AutoVap30 User Manual v 1 9 5 39 ...
Страница 6: ...TruSteel AutoVap30 User Manual v 1 9 6 39 ...
Страница 7: ...TruSteel AutoVap30 User Manual v 1 9 7 39 ...
Страница 10: ...TruSteel AutoVap30 User Manual v 1 9 10 39 Pull the Red EMERGENCY button OUT Turn all H O A switches to AUTO ...
Страница 36: ...TruSteel AutoVap30 User Manual v 1 9 36 39 AV30 FLOW IMAGE CUTAWAY CLICK HERE FOR LARGER VIEW ...
Страница 37: ...TruSteel AutoVap30 User Manual v 1 9 37 39 ...