PRIME FEED PUMP
PRIME FEED PUMP USING VACUUM
This step is critical as it will save your feed pump from running dry. Follow this step every time
you begin to feed tincture into the system before running Auto Mode.
Note: This step is only necessary if there is no liquid in the sight glass just below the preheater.
Procedures
1. Return to the HOME SCREEN [button in top left corner of HMI]
2. Open Vacuum Ball Valve
Open the Blue Ball Valve located on the vacuum line
This will introduce vacuum into the AutoVap system (as long as the Vacuum pump is on)