TruSteel AutoVap30 User Manual | | v. 1.9
19 / 39
3.
Check and Adjust Lockout Setpoints (if necessary)
Press on the lockout number you would like to adjust
Some common setpoints (setpoints may vary depending on your material and your
custom procedure)
Vacuum PSI Setpoint: 8-10 PSI
EVAP Inlet Setpoint: 150-175°F
Condenser Inlet: 46-55°F
AUTO RECOVERY MODE
Now that all set points and lockouts are in range (See ranges in PUMP CONTROL). The AV-30 is
ready to operate.
START FEED PUMP
Procedure
1.
Verify SYSTEM MODE says "EXTRACTION MODE" (far left, lower side of HMI Home
Screen)
2.
Locate the FEED PUMP area on the HMI Home Screen.
3.
Press and Hold the START button for 1-2 seconds
4.
You should see a countdown timer which will remind you to
Open the Feed Valve
5.
The Feed Pump will slowly
“ramp” up to 35 PSI (may take 3-4 minutes)
Note: The PSI will spike above 35 PSI for a moment and then will slowly decline back
down to 35 PSI. The spike is due to pressure trapped in the system.
6.
The AutoVap PSI can be adjusted up or down to the desired setpoint from the PUMP
CONTROL page.
Содержание AutoVap30
Страница 1: ...Copyright 2020 by TruSteel LLC All Rights Reserved TruSteel AutoVap30 User Manual Version 1 9 ...
Страница 3: ...TruSteel AutoVap30 User Manual v 1 9 3 39 STEP 8 SHUT DOWN 39 ...
Страница 5: ...TruSteel AutoVap30 User Manual v 1 9 5 39 ...
Страница 6: ...TruSteel AutoVap30 User Manual v 1 9 6 39 ...
Страница 7: ...TruSteel AutoVap30 User Manual v 1 9 7 39 ...
Страница 10: ...TruSteel AutoVap30 User Manual v 1 9 10 39 Pull the Red EMERGENCY button OUT Turn all H O A switches to AUTO ...
Страница 36: ...TruSteel AutoVap30 User Manual v 1 9 36 39 AV30 FLOW IMAGE CUTAWAY CLICK HERE FOR LARGER VIEW ...
Страница 37: ...TruSteel AutoVap30 User Manual v 1 9 37 39 ...