Troyka Med Inc
IOSB Mah. 2284. Cad No: 48 Yenimahalle Ankara TURKEY
Tel:
+90 3122650096- [email protected] –
www.troykamed.com
T-290 MRI Conditional Box Camera Op. Man. Pub. 022018 Pub. No: 0 Rev:0
3.3 Technical Specification
Camera
Image Sensor:
2MP CMOS Image Sensor
Signal System:
PAL/NTSC
Effective Pixels:
1944(H)×1092(V)
Min. Illumination:
Color : 0.01 Lux @(F1.2,AGC ON)
B/W : 0.001 Lux @(F1.2,AGC ON)
Shutter Time:
1/25(1/30) s to 1/50,000 s
Lens Mount:
C/CS Mount
Auto Iris:
DC Drive
Day& Night:
IR cut filter with auto switch
Wide Dynamic Range:
120dB
Video Frame Rate:
1080p@25fps/1080p@30fps
HD Video Output:
1 Analog HD output
Synchronization:
Internal Synchronization
CVBS Output:
1Vp-p Composite Output(75Ω/BNC)
S/N Ratio:
More than 62 dB
Up the Coax:
Support
General
Operating Conditions:
-30 °C ~ 60 °C (-22 °F ~ 140 °F)
Humidity 90% or less (non-condensing)
Power Supply:
12 VDC±15%
-A: 12 VDC±15%, 24 VAC±15%
Power Consumption:
Max. 4W
-A: Max. 6W
Communication :
Up the coax
Protocol: HIKVISION-C(Coaxitron)
Dimension:
69.2×57.3×148.7mm(2.7” × 2.3” × 5.85”)
Weight:
520 g
3.4 Installation
3.4.1 Installation Diagram of MRI Video Camera in Magnet Room
A representative drawing is showing the layout plan of the MRI video camera is given in Figure 2. The camera
shown in Figure 2-1 depending on the customer requirement could be installed to any place inside MRI
room. The video signal cable and 9 pin cable shown in the Figure 2-2 and 3 must be connected to the
penetration panel. Most of the MRI scanners has BNC and 9 pin filters available on the penetration panel. If
there is no available filter on your penetration panel use the ones included in your package (Table 1- 4 and
Table 1-5).
The interface box must be placed in the cabinet room. The female 9 pin connector of the interface box must
be connected to the male connector of the 9-pin filter (Figure 2-5) of the penetration panel the video signal
cable must be connected to the BNC filter. The power cable of the interface box must be connected to the
power outlet. The 30m video signal cable of the interface box must be connected to the video processing
unit in operation room. Using VGA cable connects the video interface unit to the monitor and the system will
be ready to use.