T-290 MRI Conditional Box Camera Op. Man. Pub. 022018 Pub. No: 0 Rev:0  

 

Warning! 

Self-heating of cables due to improper cable routing! Injury to equipment and/or 
operating personnel. Cables should not be looped crossed inside MRI room. 

 
 
 

 

1.4 Warnings for MRI  

 

Caution! 

Installation of materials inside the MRI must be done with extreme caution. 

 

Caution! 

Take care that ferromagnetic materials be kept at least three meters away from the 
magnet and that no installation shall be done near the filter panel if a scan is in 
progress.  

 

Caution! 

In addition, no persons with ferromagnetic prosthetic devices, such as pacemakers or 
joints replacement, should enter the MRI suite at any time. Extreme, high magnetic 
fields inside the magnetic room have the potential to dislodge items at high velocities 
and can result in serious injury, or death.   

 

Caution! 

For questions regarding installation procedures or technical support, call Troyka Med 
Service team or contact via email at 

[email protected].

 

 

 

Caution! 

Only system components explicitly designed for use inside the MRI suite should be 
placed inside the magnet room. Components not designed for MRI use may present a 
projectile hazard and can become airborne, causing serious injury, damage or death.  

 

2 General Information  

2.1. Inspection of Delivered Goods 

Each of MRI Box camera has been thoroughly tested prior to deliver and is ready for immediate use. Upon 
receipt, please report any transportation damage or missing accessories immediately. Troyka Med can only 
accept liability for such damages if it is claimed prior to initial operation. In case of transportation damage, 
please contact the Troyka Med service department. For faster support please have ready all shipment details 

Содержание T-290

Страница 1: ...4 Cad No 48 Yenimahalle Ankara TURKEY Tel 90 3122650096 info troykamed com www troykamed com T 290 MRI Conditional Box Camera Op Man Pub 022018 Pub No 0 Rev 0 OPERATORS MANUAL T 290 MRI Box Camera MRI...

Страница 2: ...85 2016 To ensure that you obtain maximum benefit from the T 290 MRI Box Camera please take a few minutes to read the enclosed information regarding operation service and maintenance After reading thi...

Страница 3: ...able risks to the patient medical staff or other persons within the MR environment 1 2 General Safety Information and Intended Use The MRI Box camera must only be operated by personnel property traine...

Страница 4: ...he Box camera possible electric shock or fire Don t handle the power cable with wet hands High voltages present could cause lethal electric shock Warning Should any foreign substances enter the device...

Страница 5: ...elds inside the magnetic room have the potential to dislodge items at high velocities and can result in serious injury or death Caution For questions regarding installation procedures or technical sup...

Страница 6: ...se consult your Troyka Med Sales or Service office Always use the power supply cord supplied in the original shipping carton If the enclosed power cord could not be used due to a different standard in...

Страница 7: ...d to view patient in MRI scanner or overview MRI room The T 290 MRI Box Camera System is designed to be MRI conditional and is manufactured and tested to the highest standards The camera system is des...

Страница 8: ...7 2 3 5 85 Weight 520 g 3 4 Installation 3 4 1 Installation Diagram of MRI Video Camera in Magnet Room A representative drawing is showing the layout plan of the MRI video camera is given in Figure 2...

Страница 9: ...ctions before cleaning 4 1 1 Monitor Surface Dampen do not saturate a clean lint free cloth with environmentally friendly glass cleaner The cleaner contains spirit as an active substance up to 98 and...

Страница 10: ...Troyka Med Inc IOSB Mah 2284 Cad No 48 Yenimahalle Ankara TURKEY Tel 90 3122650096 info troykamed com www troykamed com T 290 MRI Conditional Box Camera Op Man Pub 022018 Pub No 0 Rev 0...

Отзывы: