T-290 MRI Conditional Box Camera Op. Man. Pub. 022018 Pub. No: 0 Rev:0
Warning!
Self-heating of cables due to improper cable routing! Injury to equipment and/or
operating personnel. Cables should not be looped crossed inside MRI room.
1.4 Warnings for MRI
Caution!
Installation of materials inside the MRI must be done with extreme caution.
Caution!
Take care that ferromagnetic materials be kept at least three meters away from the
magnet and that no installation shall be done near the filter panel if a scan is in
progress.
Caution!
In addition, no persons with ferromagnetic prosthetic devices, such as pacemakers or
joints replacement, should enter the MRI suite at any time. Extreme, high magnetic
fields inside the magnetic room have the potential to dislodge items at high velocities
and can result in serious injury, or death.
Caution!
For questions regarding installation procedures or technical support, call Troyka Med
Service team or contact via email at
Caution!
Only system components explicitly designed for use inside the MRI suite should be
placed inside the magnet room. Components not designed for MRI use may present a
projectile hazard and can become airborne, causing serious injury, damage or death.
2 General Information
2.1. Inspection of Delivered Goods
Each of MRI Box camera has been thoroughly tested prior to deliver and is ready for immediate use. Upon
receipt, please report any transportation damage or missing accessories immediately. Troyka Med can only
accept liability for such damages if it is claimed prior to initial operation. In case of transportation damage,
please contact the Troyka Med service department. For faster support please have ready all shipment details