T-290 MRI Conditional Box Camera Op. Man. Pub. 022018 Pub. No: 0 Rev:0
MRI Safety Definition for MRI as Defined by International
Standards ASTM F2503-13
MR SAFE
An item that poses no known hazards resulting from exposure to any MR
environment MR SAFE items are composed of materials that are electrically
nonconductive, non-metallic, and nonmagnetic.
MR CONDITIONAL
An item with demonstrated safety in the MR environment within defined
conditions. At a minimum, address the conditions of the static magnetic
field, the switched gradient magnetic field and the radiofrequency fields.
Additional conditions, including specific conditions of the item, may be
required.
Supplementary marking – additional information that, in association with
marking as ‘’ MR CONDITIONAL ‘’ states via additional language the
conditions in which an item can be used safely within the MR environment.
MR UNSAFE
An item with poses unacceptable risks to the patient, medical staff or other
persons within the MR environment.
1.2 General Safety Information and Intended Use
The MRI Box camera must only be operated by personnel property trained in MRI safety.
The MRI Box camera must only be operated by personnel properly trained in identifying interference
problems such as artifacts, streaks and distortions in image data. It is required that as personnel handling the
MRI Box camera are familiar with the safety instructions given in the manual and other documentation
provided to ensure safe operation of the camera and associated equipment.
The system comes with a one-year warranty on all parts and labor, and a shelf life on one year starting from
the day of its original installation. Troyka Med Inc. suggests a maintenance call every six months after the
one-year original warranty period to check up on frequently used items such as connection cables and wall
mounts. This way you will make the best use of this system for your facility.
Warning!
If any of the components becomes damaged, stop using the system immediately and
notify Troyka Med Inc. customer service for assistance. Use of broken components
can cause injury to the clinician or the patient.