T-290 MRI Conditional Box Camera Op. Man. Pub. 022018 Pub. No: 0 Rev:0
Box camera serial number and damage description. It is recommended to keep and store the original
crate/box used to ship the com camera for all future transportation needs.
2.2 Power Connection Information
The MRI Box camera operates on the line voltages 90-250V AC and 50/60 Hz. It is not necessary to double
check line voltages or change any fuses. The equipment should be near the power outlet in the technical
room of MRI and the outlet should be easily accessible.
Stabilizing circuits ensures satisfactory performance within supply variations specified. If the supply voltage
in your location is not 90-250V AC, please consult your Troyka Med Sales or Service office.
Always use the power supply cord supplied in the original shipping carton.
If the enclosed power cord could not be used due to a different standard in your country, use a power cord
that conforms to the fallowing regional standards:
United States (UL)
United KingBox (BASEC/BS)
Germany (VDE)
Switzerland (SEV)
Canada (CSA)
Japan (MITI)
In other regions, please use an AC power cord that complies with the country’s safety regulations.
In order to prevent electrical hazard, the cord set must use a three-core cable of at least 6A/0.75mm
2
, and
have an earth ground contact on both the free socket and the plug. If the main cord is damaged, use only an
original replacement cable.
3 System Overview and Installation Guide
3.1 Parts Included
1 MRI conditional Box camera
2 MRI room power cable (10m cable)
Please contact for other length requirements.
3 MRI room coaxial cable (10 m cable )
Please contact for other length requirements.
4 MRI Filter: 9 Pin D-Sub LPF with EMI gasket