Communication
TCD Electronics | Function Manual | Version: TCD_PROFI_S_EN_200326_E003
11
8.
Reinstall the transmitter cassette.
9.
Reinstall the display and display cable, incl. the O-ring. Mount the O-ring by pulling it over the
display lid.
Return the device to operation
1.
Restore power to the device.
2.
Check that the device starts up as expected. If possible watch the display during startup and observe
the device versions for consistency.
3.5.
Connecting the Profibus (CH1)
WARNING!
Warnings provide very important information for the correct usage of the equipment. Not observing the
warnings may lead to danger for the equipment and to danger for health and life of the user.
1.
Remove cap and ferrule from cable gland and slide onto cable.
2.
Push cable through open gland and cable path.
3.
Restore ferrule and tighten cap to lightly hold cable in place.
4.
Signal cable screen is folded back over outer sheath and grounded beneath cable clamp.
5.
Connect wires to terminals using wiring tool, field mount transmitter
6.
Tighten cable gland.
①
In + (B)
②
In - (A)
③
Out + (B)
④
Out - (A)
Fig. 2: Profibus DP/PA