95
−
6 F 8 A 0 9 3 4
Switch operation
Display example
Description
STEP4
FIX SET
ON
SET OK?
When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .
Pushing cancels the setting operation
and exits the setting screen.
STEP5
CUR VAL
0
4.0
mA
The screen automatically changes to the
fixed current value setting screen.
The currently set value (
4.0mA
in this
example) appears.
STEP6
CUR VAL
2
0
.0
mA
Push to move the cursor to the
digit you want to change and push
to change the number of the digit.
STEP7
CUR VAL
20.0
mA
SET OK?
When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .
Pushing cancels the setting operation
and exits the setting screen.
STEP8
PLS VAL
0
0000
PPS
The screen automatically changes to the
fixed pulse rate setting screen.
The currently set value (0 PPS in this
example) appears.
STEP9
PLS VAL
00
1
00
PPS
Push to move the cursor to the
digit you want to change and push
to change the number of the digit.
STEP10(=END)
PLS VAL
00100
PPS
SET OK?
When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .
Pushing cancels the setting operation
and exits the setting screen.
N O
O K
ESC
O K
O K
N O
ESC
N O
O K
ESC
O K
O K
N O
ESC
N O
O K
ESC
O K
N O
ESC
Содержание LF516