57
−
6 F 8 A 0 9 3 4
Note 1:The meter size display loops as shown below:
①
25mm
40mm
50mm
80mm
100mm
②
②
1inch 1.5inch 2inch 3inch 4inch
①
Note 2: When the meter size is changed, range unit and count rate will be forcefully changed as
described below. If necessary, change these parameters again.
Range unit
m/s
Count rate
If the count rate goes out of the setting range
when the meter size is changed, the count rate
will be forcefully set to zero.
Note 3: The exciting frequency setting may become inappropriate for the set value when the
meter size of the detector is changed. If the exciting frequency is the value shown
below when the meter size is changed, the exciting frequency will be forcefully
changed.
Setting meter size
(mm) (inch)
Set exciting frequency
25 1
400Hz
40 ~ 80
1.5 ~ 3
200Hz
100 4
100Hz