(2) Span setting
Setting of
8.2.7 “Custom Coefficient Setting”
and
8.2.8 “Custom Unit Setting”
is applied to the
custom coefficient and unit same as
(1)Display of flow rate
.
Example
: In the case of custom unit
[dL(deciliter)/min]
.
Custom coefficient setting
・
Set the custom coefficient in accordance with
8.2.7 “Custom Coefficient
Setting”
. Set the conversion coefficient from
m
3
/min
unit to the custom
coefficient. In the case of
dL/min
, set
10000
that is the conversion coefficient
of
dL
from
m
3
.
Custom unit setting
・
Set the custom unit in accordance with
8.2.8 “Custom Unit Setting”
. Set the
character string of unit to the custom unit. In the case of this example, set ‘
d
’
‘
L
’ ‘
/
’ ‘
m
’ ‘
i
’ ‘
n
’. The rest of 1 character is blank space because the maximum
character number of the custom unit is 7.
Span
setting
・
Select the custom unit
dL/min
to the span unit in accordance with
8.2.9 “Span
(Range)”
. When custom unit is selected, the new span value will be displayed
automatically based on the custom coefficient and custom unit.