5.2 External Device Connections and Grounding
The terminal board connections of LF516/LF546 Flowmeter are shown in Figure 5.1. Proceed with wiring
as described in
Section 5.4, “Wiring Procedure.”
Figure 5.1 External Wiring Schematic Diagram
* Use a heavy copper braid or wire (cross-sectional area
5.5 mm
2
minimum
) to ground the terminal
and make it
as short as possible
as shown in Figure 5.1 for
grounding.
Also,
Avoid a common ground
where earth current may flow. (An
independent ground
is
preferable.)
* The converter has no power switch. Install the power switch at the system side.
Be sure to use a
double-pole/single-throw (both disconnection) wiring breaker.
Power supply
Wiring breaker
(double-pole/single-throw)
Output signal cable
(CVV-S)
Current output
(4-20mAdc)
or PROFIBUS
NOTE:
To avoid 2-point grounding,
ground the shield of the output
cable basically at the receiving
side.
Digital output 1
Common for DI/DO
(100 ohm or less
ground resistance)
Digital output 2
Digital input
(20 to 30 Vdc)
Power cable (CVV)
* For a 2-core cable, L1 and
L2 only.
IV wire
5.5mm
2
or more
Ground terminal
(100 ohm or less
ground resistance)
Converter unit
(100 ohm or less ground
resistance)
NOTE:
•
When
3-core power cabel
is used, do
not ground from the external
grounding terminal.(
Perform grouding
at the receving side only
).
•
Only when a
2-core power cable
is
used, perform
grounding from the
external grounding terminal.
[Instrumentation panel: Customer]
(Receiving side)