46
−
6 F 8 A 0 9 3 4
○
Operation timeout function
If no operation is made for one minute or more while the converter is in the setting mode, the
mode automatically returns to the measurement mode unless the parameters are displayed on
the screen.
Menu screen
A:DETECTOR
1 EXC CUR
2 SIZE
3 EXC FREQ
4 FLOW DIR
5 EXIT
When no operation is made
for 1 minute, the mode
automatically returns to the
measurement mode.
Parameter check screen
EXC CUR
0.2000
A
Parameter change screen
(Parameters are being
changed.)
EXC CUR
0
.2000
A
The screen does not return
to the measurement mode.
7.3.2 Setting and Calibration
In the setting mode, you can select items, or check or change the setting values as described below.
When you push or to scroll up or down the numeric value or alphabet, or when push
to move the digit, you can execute continuous operation by holding down the relevant
switch longer. (Holding down the switch longer automatically executes the operation continuously.)
●
Moving to the menu screen
Switch operation
Display example
Description
PUSH SW
CNT:
CNT CTRL
SET:
SET MODE
Mode change screen
MENU SEL
BASIC
DETAILED
PREVIEW
ZERO ADJ
PLS SET
Pushing in the measurement mode
takes you to the menu configuration
selection screen.
For configuration, select
BASIC
or
DETAILED
menu.
ESC
ESC
CNT
SET
SET
SET
ESC
ESC