56
−
6 F 8 A 0 9 3 4
Switch operation
Display example
Description
STEP6
EXC CUR
0.2150
A
SET OK?
Pushing shows a message to confirm
the setting is
OK
or not.
If
OK
, push . If you need to redo the
setting, push .
Pushing cancels the setting operation
and exits the setting screen.
STEP7(=END)
EXC CUR
0.2150
A
When you push to “
SET OK?
”
message, the set value is fixed.
Then pushing brings you back to the
setting menu. If you push , you can
change the current value from
0.2150A
again.
Note: The setting range of excitation current value is from
0.0000A
to
0.3200A
.
If you try to set an exciting current value larger than
0.3200A
, an error message appears
and the setting value returns to the previous value.
8.2.3 Meter Size
Proceed as follows to check the meter size of the detector.
Switch operation
Display example
Description
A:DETECTOR
1 EXC CUR
2 SIZE
3 EXC FREQ
4 FLOW DIR
5 EXIT
Select "
SIZE
" from the setting item
selection menu.
STEP1
SIZE
50
mm
The currently set meter size (
50mm
in this
example) appears.
Then push .
Pushing returns you to the setting
menu.
N O
O K
ESC
O K
N O
ESC
O K
O K
ESC
ESC
ESC
ESC
ESC