4. Communication With a PC
82
4.2.24 WGmKn Command
Writes the area correction factor to the instrument.
WG[m]K[n]
: m: Number of the area correction factor group. 1 to 10
n: area Number 1 to 5
PC
Command
BM-5AC
WG1K3
‗
####
‗
####
‗
####+(Cr
・
Lf)
OK+(Cr
・
Lf)
####
‗
####
‗
####+(Cr
・
Lf)
END+(Cr
・
Lf)
The values for correction factors KX, KY and KZ are entered in order to ####
‗
####
‗
####.
4.2.25 WGmLn Command
Writes the chromaticity area in the area correction factor to the instrument.
WG[m]L[n]
: m: Number of the area correction factor group. 1 to 10
n: area Number 1 to 5
PC
Command
BM-5AC
WG1L3
‗
####
‗
####
‗
####
‗
####
‗
####+(Cr
・
Lf)
OK+(Cr
・
Lf)
####
‗
####
‗
####
‗
####
‗
#### +(Cr
・
Lf)
END+(Cr
・
Lf)
The minimum value of chromaticity x, minimum value of chromaticity y, maximum value of
chromaticity x, maximum value of chromaticity y, and luminance are entered in order to
####
‗
####
‗
####
‗
####
‗
####.
4.2.26 WHO/VER/SRL Command
Reads the model name, manufacturing serial number, and program version of the instrument.
PC
command
BM-5AC
WHO/VER/SRL+(Cr
・
Lf)
OK+(Cr
・
Lf)
Model name/manufacturing serial
number/version
+(Cr
・
Lf)
END+(Cr
・
Lf)
4.2.27 CT Command
Reads the amount of time elapsed since calibration of the instrument.
PC
command
BM-5AC
CT+(Cr
・
Lf)
OK+(Cr
・
Lf)
Time [m(Cr
・
Lf)
END+(Cr
・
Lf)