3. Various Settings
58
3.2.5 Area Correction Factor
"Area correction factor" is a function for correcting measurement results that fall within an arbitrary
area when that area is specified on the xy and u'v' chromaticity diagrams.
The number of the currently set area correction factor group is displayed.
Display method
☞
"3.1.2 Displaying Data/Setting Items"
Up to 10 area correction factor groups can be set for the instrument.
Groups 1 to 5 are the area correction factors for the xy chromaticity diagram, and groups 6 to 10
are the area correction factors of the u'v' chromaticity diagram.
Also, 5 chromaticity areas can be set to 1 group.
When setting chromaticity areas, register the minimum and maximum values of chromaticity xy (or
u'v'), the minimum value of luminance L, and correction factors .
・
There are two ways of registering area correction factors, by using the CS-900A application
software (sold separately) or by making a communication program by referring to the list of
communication commands.
List of commands
☞
"4. Communication with a PC"
Memo
Area correction groups
G1
Chromaticity
area 1
G2
Chromaticity
area 2
Chromaticity
area 3
Chromaticity
area 4
G5
Chromaticity
area 5
G6
G7
G10
x,y
u', v'
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
0
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8
(y)
(x)
Specified
chromaticity area
*
C L
F A C T O R
L
G R O U P
E + 0 0
*
P l S E T L K Y = C 1 . 0 0 0 E + 0 0 *
t h I S E L K Z
O F F
. 0 0 0 E + 0 0 *
C = D E L C R = N E X T 1 S = P A G E 2