1. Before Use
19
■
Turning the Power ON/OFF
Request
Warming Up Recommendation
・
Be sure to warm up the instrument for at least 30 minutes so that measurement
can be performed at high precision.
・
After warming up has ended, be sure to press the CALIB switch to execute
calibration.
・
Turning the power ON
Flip the power switch to the left to turn the power ON.
The top screen and calibration execution screen are displayed followed by the measurement screen.
Top screen
Calibration execution screen
Measurement screen
F 2 . 0 A B S K 0 0 G 0 0 - 0
x = 0 . 4 4 7 6
y = 0 . 4 0 7 4
L = 1 . 0 0 0 E + 0 2 c d / m ^ 2
C a l i b r a t i n g
* * S t a r t B M - 5 A C * *
* *
* * V e r * . * * * *