15
7. Operation examples - Inspection mode operation
7.1.
Counter display mode without judgment
Captures, stores and outputs the peak torque (Default)
i.
Set measurement mode to “MODE-M” (Default)
ii.
Set display mode to “MEMCNT” and Hi/ Lo limits to “0” (Default)
iii.
Make sure the LCD left side display shows counter between “001” to “999” then measure
NOTE: Display can captures from about 7.5% of torque capability
iv.
When press
key after measurement, tool stores the peak torque and timestamp. If
there is wireless LAN connection on the tool, it outputs them.
NOTE: The time of timestamp is at the
key pressed. If there is Auto memory/ reset
setting the data are stored/ output automatically.
v.
Peak torque on the display will be reset and goes to next counter
NOTE: if there is a measured data on the counter number display shows it
vi.
If load the torque and exceed the value on display the peak reading is updated.
vii.
When press
key after measurement, tool stores the peak torque and timestamp. If
there is wireless LAN connection on the tool, it outputs them.
viii.
Peak torque on the display will be reset and goes to next counter