172-65639M-01 (EF200) 13 Apr 2018
69
10.3 Overview of diagnostic information
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
004
Sensor defective
1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
F
Alarm
022
Temperature
sensor defective
1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
F
Alarm
1)
046
Sensor limit
exceeded
1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
S
Warning
062
Sensor connection
defective
1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
F
Alarm
082
Data storage
1. Change main electronic
module
2. Change sensor
F
Alarm
083
Memory content
1. Restart device
2. Restore S-Dat data
3. Change sensor
F
Alarm
114
Sensor leaky
Change DSC sensor
F
Alarm
122
Temperature
sensor defective
1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
M
Warning
1)
Diagnostic of electronic
242
Software
incompatible
1. Check software
2. Flash or change main
electronics module
F
Alarm
252
Modules
incompatible
1. Check electronic modules
2. Change I/O or main electronic
module
F
Alarm
261
Electronic
modules
1. Restart device
2. Check electronic modules
3. Change I/O module or main
electronics
F
Alarm
262
Module
connection
1. Check module connections
2. Change electronic modules
F
Alarm
270
Main electronic
failure
Change main electronic module
F
Alarm
271
Main electronic
failure
1. Restart device
2. Check main electronic module
F
Alarm
272
Main electronic
failure
1. Restart device
2. Contact TLV
F
Alarm
273
Main electronic
failure
1. Emergency operation via
display
2. Change main electronics
F
Alarm
275
I/O module failure
Change I/O module
F
Alarm
276
I/O module failure
1. Restart device
2. Change I/O module
F
Alarm
277
Electronic
defective
1. Change pre-amplifier
2. Change main electronic module
F
Alarm
282
Data storage
1. Restart device
2. Contact TLV
F
Alarm
283
Memory content
1. Transfer data or reset device
2. Contact TLV
F
Alarm
302
Device verification
active
Device verification active, please
wait.
C
Warning
311
Electronic failure
1. Transfer data or reset device
2. Contact TLV
F
Alarm