172-65639M-01 (EF200) 13 Apr 2018
59
Header text
Enter display header text.
Factory setting: ------------
Separator
Select decimal separator for displaying numerical values.
• . (Period/Full stop)
• , (Comma)
Factory setting: . (Period/Full stop)
8.2.18 Configuration backup display:
Setup
→
Advanced setup
→
Configuration backup display
NOTE: Depending on operating conditions, some parameters may not be
displayed.
Descriptions of functions: Configuration backup display
Operating time
Indicates how long the device has been in operation.
Display format: Days (d), hours (h), minutes (m), seconds (s)
Last backup
Indicates when the last data backup was saved to the display
module.
Display format: Days (d), hours (h), minutes (m), seconds (s)
Configuration
management
Select action for managing the device data in the display
module.
NOTE: It is possible to save the device settings to the display
module. This function enables the user to back-up
device settings and restore data from back-ups.
Settings may be copied to other identically configured
flowmeters.
• Cancel
• Execute backup
• Restore
• Duplicate
• Compare
• Clear backup data
Factory setting: Cancel
Comparison result
Comparison between present device data and display backup.
• Settings identical
• Settings not identical
• No backup available
• Backup settings corrupt
• Check not done
• Dataset incompatible
Factory setting: Check not done